Viamed's Compatible for Draeger Evita 4 Oxygen equipment

In Stock

Main OEM

Dräger

OEM Sensor Part No.

6850645

Output Voltage

(24 - 32 μA) in air at 25°C sea level

Measurement range

0 to 100[%] O2

Output Connector

Slip rings, x 3

Accuracy

Within +/- 1[%] of full scale at constant temperature and pressure

Response time

Less than 12 seconds to 90[%] of final value

Operating humidity

0 to 99[%] RH (non-condensing)

Operating temperature

0°C to 40°C (32°F to 104°F)

Storage temperature

0°C to 50°C (32°F to 122°F)

Interference

Less than 0.5[%] O2 response to:

10.0[%] CO2 balance N2

80.0[%] N2 O balance N2

7.5[%] Halothane balance N2

7.5[%] Isoflurane balance N2

7.5[%] Enflurane balance N2

9.0[%] Sevoflurane balance N2

20.0[%] Desflurane balance N2

Expected Life

33 months in air

Warranty

15 months against defects in materials and workmanship